626B01TAE MKS Type 626B Baratron Precision 1 Torr Capacitance Manometer
Manufacturer: MKS Instruments
Product No.: 626B01TAE
Condition: In stock
Product Type: Vacuum Capacitance Manometer
Product Origin: USA
Payment: T/T, Western Union
Weight: 500g
Shipping port: Xiamen
Warranty: 12 months
Estimate delivery times: 12-26 days (International), 3-6 days (United States).
Return within 30 days of purchase. Duties & taxes are non-refundable.
Guaranteed Safe Checkout
MKS 626B01TAE | Type 626B Baratron 1 Torr Capacitance Manometer
The MKS 626B01TAE is an absolute pressure transducer from the MKS Type 626B Baratron series. It uses capacitance diaphragm technology to measure vacuum pressure independently of gas composition, within its specified operating conditions.
This configuration provides a 1 Torr full-scale range and a 15-pin male D-subminiature electrical connector. Its analog output delivers 0 to +10 VDC into a load of at least 10 kΩ.
The sensor supports low-pressure process monitoring and vacuum control applications. Its specified accuracy is ±0.25% of reading, including non-linearity, hysteresis and non-repeatability.
01 | Configuration Breakdown — Decode the Ordering Code
Ordering Code: 626B01TAE
Use the following breakdown to identify the main configuration elements. The code interpretation below follows the supplied ordering information.
-
626B
- Product family: Type 626B Baratron
- Electrical connection: 15-pin male D-subminiature connector
-
01T
- Full-scale pressure range: 1 Torr
- Range code: 01T
-
A
- Fitting configuration: 1/2-inch (12.7 mm) tubulation
- Fittings code: A
-
E
- Accuracy code: E
- Accuracy: ±0.25% of reading
- Includes non-linearity, hysteresis and non-repeatability
Ordering identification: 626B01TAE combines the Type 626B platform, 1 Torr range, 1/2-inch tubulation and accuracy code E. Confirm the complete code against the instrument nameplate before ordering.
02 | Measurement Performance and Electrical Specifications
| Specification | 626B01TAE Details |
|---|---|
| Manufacturer | MKS Instruments |
| Product family | Type 626B Baratron |
| Model | 626B01TAE |
| Product type | Capacitance diaphragm gauge / absolute pressure transducer |
| Full-scale range | 1 Torr |
| Output signal | 0 to +10 VDC |
| Minimum output load | 10 kΩ |
| Usable measurement range | 1 × 10⁻⁴ of full scale |
| Accuracy | ±0.25% of reading |
| Time constant | Less than 20 ms |
| Input power | ±15 VDC (±5%) |
| Power consumption | 35 mA |
| Electrical connector | 15-pin male D-subminiature |
| Overpressure limit | 45 psia (310 kPa), without damage |
| Ambient operating temperature | 0 to 50°C (32 to 122°F) |
| Warm-up time | 15 minutes |
03 | Low-Pressure Measurement and Temperature Stability
| Parameter | Specified Value |
|---|---|
| Lowest suggested pressure for reading | 5 × 10⁻⁴ Torr |
| Lowest suggested pressure for control | 5 × 10⁻³ Torr |
| Zero temperature coefficient | 0.015% F.S./°C |
| Span temperature coefficient | 0.04% of reading/°C |
| Internal volume, Px side | 6.3 cc |
The usable measurement range and suggested operating limits serve different purposes. The 1 × 10⁻⁴ F.S. specification describes the usable measurement range. The suggested pressure limits provide more conservative guidance for reading and process control.
For stable low-pressure operation, allow the specified 15-minute warm-up period. Avoid comparing measurements before the instrument reaches thermal stability.
04 | Mechanical Construction and Wetted Materials
- Fitting style: 1/2-inch (12.7 mm) tubulation.
- Wetted materials: Inconel and Incoloy.
- Optional fitting materials: Some configurations use 300-series stainless steel.
- Pressure sensing principle: Capacitance diaphragm.
- Pressure reference: Absolute pressure measurement.
Wetted material selection matters when the gauge contacts corrosive or process-specific gases. Confirm the exact material configuration before installation. Do not assume every optional fitting uses the same alloy as the standard sensor assembly.
05 | Installation, Cabling and Commissioning
- Verify the connection: Confirm the 15-pin male D-subminiature connector and the mating cable configuration.
- Check the supply: Provide ±15 VDC within the specified ±5% tolerance.
- Prepare the output circuit: Connect the 0 to +10 VDC output to an input with at least 10 kΩ impedance.
- Install the cable correctly: Use an overall metal-braided shielded cable.
- Ground the shield: Ground the cable shield at both ends as specified.
- Allow warm-up: Wait at least 15 minutes before evaluating measurement stability.
- Check pressure conditions: Confirm the process pressure stays within the sensor's permitted operating conditions.
Important cabling note: Shield termination is part of the measurement installation. Incorrect shielding or grounding can introduce electrical noise into the analog signal.
06 | Engineering Applications and Selection Considerations
The 1 Torr configuration suits vacuum systems that require sensitive pressure measurement over a low-pressure range. Typical application areas include:
- Vacuum chamber pressure monitoring.
- Semiconductor process equipment.
- Vacuum deposition and thin-film systems.
- Laboratory vacuum instrumentation.
- Process pressure feedback and vacuum control loops.
When selecting a replacement, compare the full-scale range, fitting, electrical connector, output signal, supply voltage and accuracy. A matching series name alone does not establish electrical or mechanical interchangeability.
07 | Related Model Comparison: 626B01TAE vs. 626B52TEE
The following comparison helps engineers distinguish the 1 Torr configuration from the 500 Torr model. The linked model is a separate product listing.
| Selection Item | 626B01TAE | 626B52TEE |
|---|---|---|
| Series | MKS Type 626B Baratron | MKS Type 626B Baratron |
| Full-scale range | 1 Torr | 500 Torr |
| Pressure application | Low-pressure vacuum measurement | Higher-pressure vacuum and process measurement |
| Range selection | Suitable when the process operates near the low-pressure end | Suitable when the process requires a substantially higher full-scale range |
| Model reference | 626B01TAE | 626B52TEE — MKS 500 Torr Baratron |
Technical selection advice: Choose the 1 Torr model when the target pressure is low and measurement resolution is the priority. Choose the 500 Torr model when the process pressure requires a wider measurement range.
Do not select a 500 Torr gauge simply because its range is larger. A higher full-scale range can reduce the useful signal span for low-pressure measurements. Conversely, a 1 Torr gauge may not suit a process that regularly approaches or exceeds its measurement range.
Before replacing either model, verify the complete ordering code, connector pinout, fitting dimensions, output scaling, power requirements and wetted materials. Confirm the 626B52TEE's detailed configuration against its product documentation rather than assuming every suffix option matches the 626B01TAE.
08 | Product Quality, Warranty and Shipping
- Warranty: 12-month warranty.
- Shipping methods: FedEx, UPS and DHL.
- Product identification: Confirm the complete model and ordering code before dispatch.
- Technical support: Contact us for configuration checks and replacement assistance.
09 | Frequently Asked Questions
Q1: Why does a 1 Torr gauge need a separate pressure limit for control applications?
Control loops need dependable feedback, not merely a detectable signal. The suggested control threshold is 5 × 10⁻³ Torr, which is higher than the suggested reading threshold. Use this distinction when defining process control limits.
Q2: Can a PLC analog input read the sensor's output directly?
Yes, if the input accepts a 0 to +10 VDC signal and presents a load impedance of at least 10 kΩ. Check the PLC input range, grounding arrangement and signal scaling before commissioning.
Q3: What happens if the cable shield connects at only one end?
The specified installation requires an overall metal-braided shield grounded at both ends. A different termination arrangement may increase noise susceptibility or create grounding problems. Follow the instrument's installation instructions and evaluate the complete grounding system.
Q4: When should engineers investigate a drifting zero reading?
First check warm-up time, ambient temperature changes, supply voltage and cable shielding. Then inspect the pressure connection for leaks or contamination. Compare the observed drift with the specified zero temperature coefficient before suspecting sensor failure.
Q5: What information should buyers provide when sourcing a replacement Baratron gauge?
Provide the complete model code, full-scale range, fitting type, connector type, output signal and supply voltage. Also confirm wetted materials and process gas compatibility. These details reduce the risk of ordering a mechanically fitting but electrically unsuitable replacement.
10 | Ordering and Technical Support
Request the MKS 626B01TAE Type 626B Baratron 1 Torr capacitance manometer by its complete ordering code. Include the equipment model and installation requirements when requesting a replacement.
Contact us for pricing, availability and technical assistance. Our team can help review the configuration and arrange shipping through FedEx, UPS or DHL.

Tags
Q: Are these parts 100% original and new?
A: Yes. Every component is guaranteed to be 100% original, brand-new, and factory-sealed. We also provide a 12-month comprehensive warranty for your peace of mind.
Q: How fast is the shipping to my country?
A: For in-stock items, we offer same-day dispatch. International express shipping (DHL/FedEx/UPS) typically delivers within 3–5 business days worldwide.
Q: Can I get a formal quotation with shipping costs?
A: Absolutely. Our sales team is ready to provide a detailed Proforma Invoice (PI) including the most competitive shipping rates for your location.
Q: Do you offer discounts for bulk orders?
A: We provide volume-based pricing for large projects or multiple-item orders. Please contact us to get our Best Price Guarantee.
12-Month Comprehensive Warranty
Every part supplied by Global PLC Spare Hub is covered by a 12-month warranty from the date of invoice.
- Coverage: Functional defects and hardware failures under normal operating conditions.
- Commitment: We provide immediate repair, replacement, or a full refund for any verified defective components to minimize your operational downtime.
Returns & Exchange Policy
We offer a 30-day return policy to ensure you receive the exact part your system requires.
- Eligibility: Returns are accepted for uninstalled items in their original, factory-sealed packaging.
- RMA Process: Please contact our support team to obtain a Return Merchandise Authorization (RMA) number before shipping.
- Fast Processing: Refunds or exchanges are processed within 3–5 business days after the returned item passes our technical inspection.
Need Technical Assistance?Not sure if this part is compatible with your system? [Consult our Engineers] before purchasing to ensure 100% compatibility.