Home 722B12TCD2FA MKS Instruments 4-VCR Precision Pressure Sensor

722B12TCD2FA MKS Instruments 4-VCR Precision Pressure Sensor

  • Manufacturer: MKS Instruments

  • Product No.: 722B12TCD2FA

  • Condition: In stock

  • Product Type: Absolute Pressure Sensor

  • Product Origin: USA

  • Payment: T/T, Western Union

  • Weight: 500g

  • Shipping port: Xiamen

  • Warranty: 12 months

Selected quantity: 1
Request a quote

Estimate delivery times: 12-26 days (International), 3-6 days (United States).

Return within 30 days of purchase. Duties & taxes are non-refundable.

Guaranteed Safe Checkout

Visa
PayPal
Mastercard
American Express

Product Overview | Compact Absolute Pressure Measurement

MKS Instruments 722B12TCD2FA is a compact absolute pressure transducer from the Baratron® 722B Series. It uses capacitance diaphragm technology to measure vacuum and process pressure independently of gas composition.

This configuration provides a 100 Torr full-scale range and ±0.5% of reading accuracy. Its process interface uses a female 4-VCR fitting. A 9-pin Type D connector handles power and analog signal connections.

The transducer accepts +13 to +32 VDC and provides a proportional 0–10 VDC output. Its all-metal wetted construction uses Inconel® for compatibility with many industrial process gases.

A response time below 20 milliseconds supports dynamic pressure monitoring. The robust sensor construction also tolerates specified pressure excursions. These characteristics make the 722B12TCD2FA suitable for vacuum equipment, semiconductor systems, and industrial process monitoring.

01 | Configuration Breakdown – Model Suffix Analysis

The complete ordering code identifies the pressure range, process fitting, electrical configuration, accuracy, and connector. Each suffix defines a specific hardware requirement.

Code Segment Configuration Technical Meaning
722B Product Series Compact general-purpose absolute Baratron® capacitance manometer
12T Pressure Range 100 Torr full scale
CD Process Fitting 1/4-inch 4-VCR female connection
2 Input / Output +13 to +32 VDC input / 0–10 VDC output
F Accuracy Grade ±0.5% of reading
A Electrical Connector 9-pin Type D connector

Complete Model: 722B12TCD2FA

Configuration Architecture:

  • Pressure Measurement: 100 Torr absolute pressure.
  • Vacuum Connection: Female 4-VCR process fitting.
  • Power Architecture: Single positive DC supply, +13 to +32 VDC.
  • Signal Architecture: 0–10 VDC proportional analog output.
  • Measurement Accuracy: ±0.5% of reading.
  • Electrical Interface: 9-pin D-subminiature connector.

Configuration Distinction: The CD suffix specifies 4-VCR female. The CE suffix specifies 8-VCR female. These fittings are not interchangeable, so verify the complete ordering code before purchasing.

02 | Technical Specification Register

Parameter Specification
Manufacturer MKS Instruments, USA
Model 722B12TCD2FA
Product Series Baratron® 722B
Product Type Capacitance Diaphragm Pressure Transducer
Measurement Type Absolute Pressure
Full-Scale Range 100 Torr
Accuracy ±0.5% of reading
Response Time Less than 20 ms
Analog Output 0–10 VDC
Minimum Output Load Greater than 10 kΩ
Power Supply +13 to +32 VDC
Maximum Supply Current 10 mA
Process Fitting 1/4-inch 4-VCR female
Electrical Connector 9-pin Type D
Wetted Material Inconel®
Wetted Surface Finish ≤10 Ra
Operating Temperature 0°C to 50°C
Zero Temperature Coefficient 0.008% F.S./°C
Span Temperature Coefficient 0.04% of reading/°C
Overpressure Limit 45 psia or 2 × F.S., whichever is greater
Burst Pressure 10 × F.S. or 100 psi, whichever is greater

03 | Sensing Technology and Measurement Characteristics

The Baratron® 722B measures pressure through capacitive changes in a precision diaphragm assembly.

Pressure acts on the sensing diaphragm and produces mechanical displacement. The internal electronics convert the resulting capacitance variation into a proportional voltage signal.

This measurement method offers several engineering advantages:

  • Gas-Independent Measurement: The sensing principle does not depend on gas thermal conductivity.
  • Absolute Pressure Reference: Measures pressure against an internal reference vacuum.
  • Fast Response: The specified response time remains below 20 milliseconds.
  • Stable Measurement: The accuracy specification includes non-linearity, hysteresis, and non-repeatability.
  • Corrosion Resistance: Inconel® wetted surfaces support compatibility with many process gases.
  • Clean Surface Construction: Wetted surfaces receive a controlled finish of ≤10 Ra.

Accuracy Definition: The ±0.5% specification refers to the pressure reading. It does not represent ±0.5% of full scale.

04 | Analog Output and Electrical Interface

The 722B12TCD2FA uses a 0–10 VDC output for integration with compatible pressure controllers and industrial analog input systems.

Electrical Parameter Specification
Supply Voltage +13 to +32 VDC
Supply Current 10 mA maximum
Output Signal 0–10 VDC
Output Load Greater than 10 kΩ
Connector Type 9-pin Type D
Signal Reference Pressure Return
Power Reference Power Return

Nominal Signal Scaling: The 100 Torr full-scale range corresponds to a 0–10 VDC output.

  • 0 Torr corresponds to approximately 0 VDC.
  • 25 Torr corresponds to approximately 2.5 VDC.
  • 50 Torr corresponds to approximately 5 VDC.
  • 75 Torr corresponds to approximately 7.5 VDC.
  • 100 Torr corresponds to approximately 10 VDC.

These values represent nominal linear scaling. Actual readings depend on calibration and instrument specifications.

05 | 9-Pin D-Sub Connector Pin Assignment

The following pinout follows the MKS 722B instruction manual. Keep the pressure signal return and power return connections distinct during cable assembly.

Pin Number Signal Assignment Function
Pin 1 Pressure Output 0–10 VDC analog pressure signal
Pin 2 No Connection Not internally connected
Pin 3 No Connection Not internally connected
Pin 4 + Power Input +13 to +32 VDC supply
Pin 5 No Connection Not internally connected
Pin 6 No Connection Not internally connected
Pin 7 No Connection Not internally connected
Pin 8 Pressure Return Analog signal reference
Pin 9 Power Return DC supply return

Wiring Precaution: Do not assume that pressure return and power return are interchangeable. Follow the MKS wiring diagram and compatible controller requirements.

Verify connector orientation before assigning pin numbers. Incorrect wiring can cause unstable readings or damage connected equipment.

06 | Pressure Limits and Thermal Performance

The sensor construction supports pressure excursions beyond its normal measurement range. Engineers must distinguish the specified overpressure limit from the burst pressure rating.

Parameter Value
Full-Scale Pressure 100 Torr
Overpressure Limit 45 psia or 2 × full scale, whichever is greater
Burst Pressure 10 × full scale or 100 psi, whichever is greater
Operating Temperature 0°C to 50°C
Zero Temperature Coefficient 0.008% F.S./°C
Span Temperature Coefficient 0.04% of reading/°C

Engineering Note: Overpressure tolerance does not authorize continuous operation above the rated measurement range. Avoid unnecessary pressure surges during system venting and pump-down.

07 | Vacuum Connection and Wetted Material

The CD fitting configuration provides a female 4-VCR process interface. It supports compatible metal-gasket vacuum connections.

  • Fitting Type: 1/4-inch 4-VCR female.
  • Wetted Material: Inconel® nickel-based alloy.
  • Surface Finish: ≤10 Ra on wetted surfaces.
  • Construction: All-metal sensor wetted surfaces.
  • Material Compatibility: Suitable for many compatible semiconductor and industrial process gases.

Some optional fitting configurations use 316 stainless steel. The standard wetted sensor surfaces use Inconel®. Confirm the actual fitting material when chemical compatibility is critical.

Installation Warning: Use compatible VCR gaskets and approved assembly procedures. Do not apply excessive tightening force to the fitting or sensor body.

08 | Installation and Commissioning Guide

Proper installation protects the sensing diaphragm and maintains vacuum system integrity. Complete the following checks before applying operating pressure.

Mechanical Installation

  • Verify the 4-VCR female connection against the process tubing.
  • Inspect the sealing surfaces before installing the gasket.
  • Use compatible metal gaskets for the process conditions.
  • Check the mounting position and avoid unnecessary mechanical stress.
  • Inspect all vacuum joints for leakage after assembly.

Electrical Installation

  • Provide a regulated +13 to +32 VDC supply.
  • Connect the positive supply to Pin 4.
  • Connect the power return to Pin 9.
  • Connect the analog output through Pin 1.
  • Connect the pressure return through Pin 8.
  • Verify the receiving input impedance exceeds 10 kΩ.

Commissioning Sequence

  1. Inspect the complete electrical and vacuum connections.
  2. Verify the power supply voltage and polarity.
  3. Allow the instrument to reach stable operating conditions.
  4. Check the analog output at a known pressure.
  5. Compare readings against a calibrated reference instrument.
  6. Configure the PLC or controller for a 0–100 Torr range.
  7. Perform a final leak and signal stability check.

09 | Maintenance and Troubleshooting

The compact sensor architecture simplifies integration into vacuum equipment. Routine checks should focus on pressure connections, power quality, and analog signal integrity.

Observed Problem Engineering Inspection
Output remains at zero Check supply voltage, Pin 4, Pin 9, and pressure conditions.
Output fluctuates unexpectedly Inspect grounding, shielding, and analog input stability.
Reading differs from reference Verify calibration, temperature, and controller scaling.
Pressure does not stabilize Inspect vacuum joints, process leakage, and pump performance.
Incorrect PLC reading Check signal return, input impedance, and 0–10 V configuration.
Sudden measurement shift Investigate pressure surges and abnormal process conditions.

Maintenance Recommendation: Inspect the VCR sealing interface whenever technicians disconnect the sensor. Replace damaged sealing components and repeat the leak test after reassembly.

10 | System Compatibility and Integration

The MKS Baratron 722B supports vacuum control systems with compatible DC power and voltage-input measurement channels.

  • PLC Systems: Use a 0–10 VDC analog input module.
  • DCS Systems: Connect through a compatible voltage signal acquisition channel.
  • Vacuum Controllers: Integrate with controllers supporting the 722B electrical interface.
  • Semiconductor Equipment: Monitor absolute pressure in compatible process chambers.
  • Vacuum Process Equipment: Provide pressure feedback for monitoring and control.

Compatibility Requirements:

  • Verify the +13 to +32 VDC supply requirement.
  • Confirm the controller accepts a 0–10 VDC signal.
  • Maintain an input impedance above 10 kΩ.
  • Match the female 4-VCR fitting with the process connection.
  • Confirm the 9-pin D-sub pinout before cable fabrication.

Do not connect the output directly to a 4–20 mA input. Use an appropriate signal converter when the existing control system lacks voltage-input capability.

11 | Typical Industrial Applications

  • Semiconductor Manufacturing: Monitor vacuum pressure in compatible process chambers.
  • Thin-Film Deposition: Provide pressure feedback during vacuum coating operations.
  • Vacuum Processing: Measure absolute pressure in industrial vacuum equipment.
  • Research Laboratories: Support controlled pressure measurement in experimental systems.
  • Vacuum Equipment Integration: Supply analog pressure signals to automation controllers.
  • Process Monitoring: Track pressure changes during production and equipment operation.

12 | Product Selection and Replacement Guidance

When selecting a replacement for an existing Baratron® transducer, compare the complete ordering code rather than the series name alone.

Selection Criteria Required Configuration
Product Series 722B
Measurement Range 100 Torr
Fitting 4-VCR female
Power Input +13 to +32 VDC
Signal Output 0–10 VDC
Accuracy ±0.5% of reading
Connector 9-pin Type D

Important Selection Distinction: The 722B12TCD2FA and 722B12TCE2FA share the 100 Torr range and electrical configuration. However, CD specifies 4-VCR female, while CE specifies 8-VCR female.

Do not substitute these configurations without checking the process connection. Also verify the power supply and controller scaling when replacing older pressure transducers.

13 | Packaging, Warranty and International Shipping

  • Product: MKS Instruments 722B12TCD2FA Baratron® Pressure Transducer.
  • Configuration: 100 Torr, 4-VCR female, 0–10 VDC output.
  • Electrical Interface: 9-pin D-sub connector.
  • Packaging: Protective packaging for precision industrial instrumentation.
  • Documentation: Product identification and available technical documentation.

Warranty: 12-month seller warranty.

Shipping Options: FedEx, UPS, and DHL international delivery.

Product Condition: New and original factory-supplied equipment.

Ordering Note: Confirm the fitting, connector, and complete model suffix before ordering. Cables, gaskets, controllers, and additional accessories require separate confirmation.

14 | Frequently Asked Questions

Q1: Can the 722B12TCD2FA measure pressure when the process gas composition changes?

Yes. Its capacitance diaphragm principle measures pressure through mechanical displacement rather than gas thermal conductivity. Changes between compatible process gases do not require thermal conductivity correction factors. Engineers must still evaluate chemical compatibility with exposed materials.

Q2: Why is the separation between pressure return and power return important?

These connections serve different electrical functions. Pin 8 provides the pressure signal reference. Pin 9 provides the power return. Incorrect grounding arrangements can introduce measurement errors or damage connected equipment. Follow the specified interface diagram during cable fabrication.

Q3: How should engineers evaluate pressure spikes during pump switching?

Monitor pressure changes during pump startup, isolation valve operation, and chamber venting. Compare transient pressure against the specified overpressure limit. The protective rating helps tolerate occasional excursions, but it does not replace proper pressure management.

Q4: Can a 100 Torr transducer replace a 250 Torr or 500 Torr model?

Not automatically. A lower full-scale range changes the pressure-to-voltage relationship and the available measurement span. Check the maximum process pressure and controller scaling before substitution. Select a range that accommodates normal operation and expected pressure excursions.

Q5: What should purchasing engineers verify when ordering a replacement for a CD or CE configuration?

Check the exact VCR fitting size, connector type, output voltage, and power requirement. CD identifies 4-VCR female, while CE identifies 8-VCR female. Request confirmation using the complete nameplate model number. This check helps prevent mechanical installation problems and unnecessary system modifications.

Precision Baratron Pressure Measurement for Vacuum Systems

The MKS Instruments 722B12TCD2FA combines 100 Torr absolute pressure measurement with ±0.5% reading accuracy and a compact capacitance diaphragm design.

Its female 4-VCR process connection, 9-pin D-sub interface, and 0–10 VDC output support integration into compatible vacuum equipment. Inconel® wetted surfaces and fast response make this configuration suitable for demanding industrial pressure monitoring applications.

Contact us for a quotation, stock availability, delivery information, or replacement verification. Provide your complete model number and application requirements for technical assistance.

Tags

Q: Are these parts 100% original and new?

A: Yes. Every component is guaranteed to be 100% original, brand-new, and factory-sealed. We also provide a 12-month comprehensive warranty for your peace of mind.

Q: How fast is the shipping to my country?

A: For in-stock items, we offer same-day dispatch. International express shipping (DHL/FedEx/UPS) typically delivers within 3–5 business days worldwide.

Q: Can I get a formal quotation with shipping costs?

A: Absolutely. Our sales team is ready to provide a detailed Proforma Invoice (PI) including the most competitive shipping rates for your location.

Q: Do you offer discounts for bulk orders?

A: We provide volume-based pricing for large projects or multiple-item orders. Please contact us to get our Best Price Guarantee.




12-Month Comprehensive Warranty

Every part supplied by Global PLC Spare Hub is covered by a 12-month warranty from the date of invoice.

  • Coverage: Functional defects and hardware failures under normal operating conditions.
  • Commitment: We provide immediate repair, replacement, or a full refund for any verified defective components to minimize your operational downtime.

Returns & Exchange Policy

We offer a 30-day return policy to ensure you receive the exact part your system requires.

  • Eligibility: Returns are accepted for uninstalled items in their original, factory-sealed packaging.
  • RMA Process: Please contact our support team to obtain a Return Merchandise Authorization (RMA) number before shipping.
  • Fast Processing: Refunds or exchanges are processed within 3–5 business days after the returned item passes our technical inspection.

Need Technical Assistance?Not sure if this part is compatible with your system? [Consult our Engineers] before purchasing to ensure 100% compatibility.