MKS 627H.1TBD2B – Heated Baratron Vacuum Pressure Transducer

In low-pressure vacuum systems, thermal stability directly affects measurement consistency. The MKS Instruments 627H.1TBD2B addresses this requirement through a temperature-controlled Baratron sensing assembly.

This model belongs to the MKS 627H Series (600 Series). It provides absolute pressure measurement with a 0.1 Torr full-scale range. Its internal sensor maintains a controlled temperature of 45°C.

The heated sensing design helps reduce measurement variations caused by changes in sensor temperature. A built-in heater status LED provides a convenient visual indication of heater operation.

With vertical calibration, an analog 0–10 VDC output, and an 8 VCR female process connection, this transducer supports precision vacuum monitoring in semiconductor and industrial process equipment.

01 | Measurement Architecture

Technical Element Specification Engineering Significance
Manufacturer MKS Instruments Industrial vacuum measurement technology
Model 627H.1TBD2B Specific ordering configuration
Product Family 627H Series Heated Baratron product family
Product Series 600 Series Vacuum pressure measurement platform
Measurement Technology Baratron Capacitance Manometer Capacitance-based pressure measurement
Measurement Type Absolute Pressure Measures pressure relative to an internal vacuum reference
Full-Scale Range 0.1 Torr Designed for low-pressure vacuum monitoring
Sensor Temperature 45°C controlled Stabilizes the sensing assembly temperature
Calibration Orientation Vertical Requires the specified calibration orientation
Warm-Up Time 4 hours Allows thermal stabilization before precision measurements
Heater Status Indicator LED Provides local heater operating status indication

02 | Measurement Performance

The 627H.1TBD2B combines a low-pressure measurement range with defined accuracy and resolution specifications.

Performance Parameter Value
Full-Scale Pressure 0.1 Torr
Measurement Accuracy 0.12% of Reading
Accuracy Components Includes non-linearity, hysteresis, and non-repeatability
Resolution 0.001% of Full Scale
Equivalent Resolution 0.000001 Torr
Zero Temperature Coefficient 0.005% FS/°C
Span Temperature Coefficient 0.02% of Reading/°C
Sensor Temperature 45°C controlled
Warm-Up Requirement 4 hours

Engineering Note: Accuracy and resolution describe different measurement characteristics. Resolution indicates the smallest specified measurement increment. It does not represent absolute measurement accuracy.

03 | Electrical Interface and Signal Integration

The transmitter provides a voltage output for compatible vacuum monitoring and process control systems.

Electrical Parameter Specification
Supply Voltage ±15 VDC, ±5%
Maximum Current 0.25 A
Analog Output 0 to +10 VDC
Output Load Greater than 10 kΩ
Electrical Connector 15-pin D-sub
Connector Retention Thread-locking hardware
Trip Point Outputs Not provided
Heater Status Visual LED indicator

Signal Integration Considerations:

  • Use a compatible ±15 VDC power supply.
  • Verify the analog input range before connecting the output.
  • Maintain an input impedance above 10 kΩ.
  • Check the 15-pin D-sub connector pin assignments.
  • Provide suitable cable shielding and grounding practices.
  • Do not configure external trip functions through this analog-only output.

04 | Process Connection and Mechanical Design

The process interface and wetted construction support integration into clean vacuum systems.

Mechanical Parameter Specification
Process Fitting 8 VCR Female
Wetted Material Inconel
Internal Volume 6.3 cc
Overpressure Limit 45 psia (310 kPa) or 120% F.S., whichever is greater
Body Diameter 3.15 in (80.0 mm)
Body Height 3.87 in (98.4 mm)
Stem Length 1.38 in (35.1 mm)
Compliance CE compliant

Mechanical Selection Note: Confirm the 8 VCR female connection before ordering. Check the available installation clearance against the 80.0 mm body diameter and 98.4 mm body height.

05 | Thermal Management and Operating Environment

The heated sensor maintains its sensing temperature at 45°C. This design helps control temperature-related measurement drift.

Environmental Parameter Specification
Operating Temperature 15 to 40°C
Controlled Sensor Temperature 45°C
Zero Temperature Coefficient 0.005% FS/°C
Span Temperature Coefficient 0.02% of Reading/°C
Heater Indication LED
Thermal Warm-Up 4 hours

Important: The 45°C value describes the controlled sensor temperature. The specified ambient operating range remains 15–40°C.

06 | Installation and Commissioning Procedure

Proper installation preserves the calibration conditions and protects the low-pressure sensing assembly.

  1. Verify Orientation: Install the instrument according to its vertical calibration specification.
  2. Inspect the Process Port: Check the 8 VCR female fitting for contamination or mechanical damage.
  3. Prepare the Vacuum Line: Ensure the connection surfaces meet the process cleanliness requirements.
  4. Secure the Connection: Use compatible VCR hardware and approved installation procedures.
  5. Connect Power: Apply the specified ±15 VDC supply within its tolerance.
  6. Wire the Output: Connect the 0–10 VDC signal to a compatible analog input.
  7. Verify Loading: Maintain an output load above 10 kΩ.
  8. Allow Warm-Up: Wait four hours before relying on precision readings.
  9. Check Heater Status: Inspect the heater status LED during startup.
  10. Validate the Reading: Compare the output with an appropriate calibrated pressure reference.

07 | Maintenance and Troubleshooting Insights

  • Unstable Readings: Check the warm-up period, power stability, grounding, and analog input loading.
  • Zero Offset: Verify the reference pressure and calibration conditions before adjusting the system.
  • Slow Stabilization: Confirm the heater status and allow the complete four-hour warm-up.
  • Unexpected Pressure Values: Inspect the process connection and verify the instrument's measurement range.
  • Output Problems: Check the ±15 VDC supply and 15-pin connector wiring.
  • Installation-Related Drift: Confirm the vertical calibration orientation and mounting arrangement.

Maintenance Recommendation: Avoid opening the sensing assembly or changing internal calibration settings without authorized procedures. Use suitable reference equipment for periodic verification.

08 | Application and System Compatibility

The MKS 627H.1TBD2B suits vacuum applications that require low-pressure absolute measurement and a heated sensing assembly.

  • Semiconductor process vacuum monitoring
  • Thin-film deposition equipment
  • Vacuum chamber pressure measurement
  • Precision gas process monitoring
  • Industrial vacuum instrumentation
  • Vacuum control and data acquisition systems

System Integration Checklist:

  • Analog input must accept 0–10 VDC.
  • Power supply must provide ±15 VDC.
  • Analog input impedance must exceed 10 kΩ.
  • Mechanical connections must match 8 VCR female fittings.
  • Installation must respect the vertical calibration condition.
  • Ambient temperature must remain within 15–40°C.

09 | Product Compliance and Documentation

  • CE Compliance: The specified model carries CE compliance.
  • Electrical Configuration: ±15 VDC power and 0–10 VDC analog output.
  • Process Materials: Inconel wetted construction.
  • Calibration Configuration: Vertical calibration.

Request the applicable product declaration and supporting technical documents when your project requires formal compliance verification.

10 | Five Engineering FAQs

Q1: Why does a heated Baratron use a 45°C sensor temperature instead of ambient temperature?

A: Temperature changes can affect the mechanical and electrical characteristics of a capacitance sensing assembly. Controlled heating helps maintain consistent thermal conditions and reduces temperature-related measurement variations.

Q2: How should engineers scale the 0–10 VDC output in a PLC or data acquisition system?

A: Use the instrument's 0.1 Torr full-scale range as the scaling reference. A linear 0–10 V mapping corresponds to 0–0.1 Torr. Confirm the transfer function and polarity against the applicable manufacturer documentation before commissioning.

Q3: What happens if the gauge operates before completing its four-hour warm-up?

A: The sensor may not reach its specified thermal equilibrium. Early readings may therefore differ from stabilized measurements. Allow the full warm-up period before conducting calibration checks or process qualification.

Q4: Why is vertical calibration important when replacing an existing vacuum transducer?

A: Installation orientation can affect the relationship between the sensor's calibrated condition and its installed operating condition. A replacement should preserve the specified vertical orientation whenever practical. Verify zero performance after installation.

Q5: How can engineers distinguish a gauge fault from a vacuum system problem?

A: Compare the gauge output with an independent calibrated reference. Check the power rails, analog input loading, warm-up status, and process connection. If both instruments show similar pressure changes, investigate the vacuum system before replacing the transducer.

11 | Ordering and Supply Information

For accurate model selection, provide the complete ordering number and installation requirements when requesting a quotation.

Ordering Information Details
Manufacturer MKS Instruments
Part Number 627H.1TBD2B
Series 627H / 600 Series
Product Type Heated Baratron Absolute Pressure Transducer
Full-Scale Range 0.1 Torr
Output 0–10 VDC
Process Connection 8 VCR Female
Warranty 12 months
Shipping Options FedEx, DHL, UPS

Contact us for a quotation, delivery information, and technical assistance. Our team can help verify the model, electrical interface, pressure range, and process connection before ordering.

Tags

Q: Are these parts 100% original and new?

A: Yes. Every component is guaranteed to be 100% original, brand-new, and factory-sealed. We also provide a 12-month comprehensive warranty for your peace of mind.

Q: How fast is the shipping to my country?

A: For in-stock items, we offer same-day dispatch. International express shipping (DHL/FedEx/UPS) typically delivers within 3–5 business days worldwide.

Q: Can I get a formal quotation with shipping costs?

A: Absolutely. Our sales team is ready to provide a detailed Proforma Invoice (PI) including the most competitive shipping rates for your location.

Q: Do you offer discounts for bulk orders?

A: We provide volume-based pricing for large projects or multiple-item orders. Please contact us to get our Best Price Guarantee.




12-Month Comprehensive Warranty

Every part supplied by Global PLC Spare Hub is covered by a 12-month warranty from the date of invoice.

  • Coverage: Functional defects and hardware failures under normal operating conditions.
  • Commitment: We provide immediate repair, replacement, or a full refund for any verified defective components to minimize your operational downtime.

Returns & Exchange Policy

We offer a 30-day return policy to ensure you receive the exact part your system requires.

  • Eligibility: Returns are accepted for uninstalled items in their original, factory-sealed packaging.
  • RMA Process: Please contact our support team to obtain a Return Merchandise Authorization (RMA) number before shipping.
  • Fast Processing: Refunds or exchanges are processed within 3–5 business days after the returned item passes our technical inspection.

Need Technical Assistance?Not sure if this part is compatible with your system? [Consult our Engineers] before purchasing to ensure 100% compatibility.